Chong Chen and Hui Li
Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Science, Suzhou, 215163, China
Life was assembled from molecules, subcellular organelles, cells, tissues, organs, to the whole organism in totally different ways. The assembly at each level has its own structure, dynamics, and functions, making for such a complex and beautiful living world. To study such complex structures, as nobel prize winner Feynmann said, "It is very easy to understand many of these fundamental biology questions: you just look at the things." However, different imaging tools need to be developed for different purposes to look at the various biological objects, with a scale from nanometers to centimeters.
Optical microscopy plays the most important role in inspecting the microscale biological world among all the imaging tools. Although optical microscopy has been invented for more than 300 years, we have witnessed significant improvement in the optical microscope technique in the last 30 years. These improvements mostly fall into two aspects: sample labeling technique and different imaging modalities.
Image contrast is the first factor to be concerned with for optical imaging. By now, fluorescent imaging has provided the highest contrast due to the filtering out of excitation light. Organic dyes, quantum dots, and fluorescent protein are the three most widely used fluorescent labeling agents. Fluorescent proteins, which won the novel prize for chemistry in 2008, provide a genetic way for labeling so that fluorescent imaging with live cells, organelles, and even live animals becomes possible.
High-end microscopes fall into two categories: widefield microscopes and point scanning microscopes (Figure 1.1). A wide-field microscope takes imaging by a camera and usually has high speed and high photon efficiency. Typical examples are the Tirf-microscope, structured illumination microscope, and single-molecule localization super-resolution microscope. The point-scanning microscope takes imaging by fast scanning the excitation laser beam or sample, usually at a lower speed but with higher axial sectioning capability. Typical examples include laser scanning confocal microscopes, two-photon microscopes, and STED super-resolution.
Figure 1.1 (a) Comparison of wide-field microscopy. (b) Point-scanning microscopy.
Source: Chong Chen.
This chapter mainly introduces the advance of the wide-field fluorescent microscope in the last ten years. We first introduce the methods to improve the optical sectioning and the resolution by structured illumination, then introduce the methods by light sheet illumination. The optical principle, setup, image processing method will be introduced in each section. The chapter ends with a prospect for future development.
The optical section in microscopy defines its capability to resolve structure axially. In an epi-fluorescent microscope, the entire sample space is illuminated, and all of the excited fluorescence signals collected by the objective can go into the array detector. Consequently, when the sample goes out of focus, its image becomes blurred, but the signals do not disappear. This problem presents a significant hindrance in wide-field microscopy.
In optical microscopy, the depth of focus is how far the sample plane can move while the specimen remains in perfect focus. The numerical aperture of the objective lens is the main factor that determines the depth of focus (D):
where ? is the wavelength of the fluorescent light, n is the refractive index of the medium [usually air (1.000) or immersion oil (1.515)], NA is the numerical aperture of the objective lens. The variable e is the smallest distance that can be resolved by a detector that is placed in the image plane of the microscope objective, whose lateral magnification is M. For a high-end fluorescent microscope with NA 1.4 and a 100× magnification objective, the depth of focus is on the order of 500 to 700?nm, dependent on the fluorescent wavelength. This depth of field defines the best optical section capability for an epi-fluorescence microscope.
Figure 1.2 (a) Sketch of the depth of field. (b) The defocused signals cause blur of the focused image.
However, when imaging samples with a thickness larger than the microscope's depth-of-focus, the sample's out of focus plane is also excited and forms a defocus image at the camera plane (Figure 1.2). The superimposition of these defocus images lowers the contrast of the camera's captured image and practically lowers the axial resolution. The imperfections in the microscope's optics and the scattering of fluorescence signals by the sample itself make the situation even worse. So, the priority demand to improve the wide-field microscope's performance lies in eliminating the out-of-focus signal, yielding better optical sectioning capability.
In a laser scanning confocal microscope, the out-of-focus light is rejected using a pinhole. In a wide-field microscope, no pinhole can be used since it uses a array detector. One way to reject the out-of-focus signals is by using structured illumination. By utilizing a grating or a digital mirror device (DMD), stripe patterns were projected on the image plane so that a structured illumination was created to excite the fluorescence molecules within the focus plane. The actual focus range of the stripes can be made very sharp if the proper period of the grating is used. Out of focus, the strip patterns become uniform, which will generate a nonmodulated background. Therefore, the image formed by the microscope will consist of striped in-focus features superposed with uniformly illuminated out-of-focus features. A postprocessing algorithm could reject this background afterward, thus obtaining a better optical section capability.
So, to obtain optical sectioning with structured illumination, two factors are needed: optical instrumentation to create structured illumination and the optical section reconstruction algorithm. These two aspects will be discussed in the following section.
For structured illumination, the excitation light field needs to be patterned, and the pattern needs to be shifted or rotated to capture all sample information. Several methods have been developed for this purpose.
Figure 1.3 Structured illumination with grating for optical section.
Figure 1.4 Structured illumination with digital mirror device.
Dateiformat: ePUBKopierschutz: Adobe-DRM (Digital Rights Management)
Systemvoraussetzungen:
Das Dateiformat ePUB ist sehr gut für Romane und Sachbücher geeignet – also für „fließenden” Text ohne komplexes Layout. Bei E-Readern oder Smartphones passt sich der Zeilen- und Seitenumbruch automatisch den kleinen Displays an. Mit Adobe-DRM wird hier ein „harter” Kopierschutz verwendet. Wenn die notwendigen Voraussetzungen nicht vorliegen, können Sie das E-Book leider nicht öffnen. Daher müssen Sie bereits vor dem Download Ihre Lese-Hardware vorbereiten.Bitte beachten Sie: Wir empfehlen Ihnen unbedingt nach Installation der Lese-Software diese mit Ihrer persönlichen Adobe-ID zu autorisieren!